Skip to main content

Table 4 Method validation parameters for the quantification of colchicine

From: Screening the elite chemotypes of Gloriosa superba L. in India for the production of anticancer colchicine: simultaneous microwave-assisted extraction and HPTLC studies

Sr.no.

Parameters

Colchicine

1

Instrument precision [%CV, n=10]

0.44

2

Repeatability [%CV, n=10]

0.58

3

Linearity range (ng/spot; n=12a

1000–5000

4

Correlation coefficient (r2)

0.975

5

Regression equation (Linear)

Y= 9178+ 4.527*X

6

Calculated SD value (CATS software)

10.27

7

bLimit of detection (LOD) (ng) [3 x SD/S] (ng/band)

70

8

bLimit of quantitation (LOQ) (ng) [10 x SD/S] (ng/band)

210

9

Rf and λmax

0.27; 254 nm

  1. aFour concentration levels in triplicate; bSD is the standard deviation of the blank response and S is the slope of the calibration plot